Search result
"Japanese Pharmacopoeia" Filtered 14 Item.
GC
- Samples
- Ethylene glycol , Diethylene glycol
- Applications
- Analysis of macrogol 400 using Shinwasorb
- Column
- Sorbitol 3% Shinwasorb-S 80/100
- Column Size
- 1.6 m x 3.2 mm I.D. Glass
- Column temp.
- 160℃
- Carrier gas
- 47 mL/min N2
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 2.0 μL
View detail
GC
- Samples
- Cyclohexane , N,N-dimethylaniline , Naphthalene ,
- Applications
- Analysis of ampicillin hydrate using Shinwasorb
- Column
- Silicone OV-17 1.5% Shinwasorb-S 60/80
- Column Size
- 2.1 m x 2.6 mm I.D. Glass
- Column temp.
- 105℃
- Carrier gas
- 20 mL/min He(adjust so that the retention time of N,N-dimethylaniline is about 5 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Cyclohexane , Ethyl acetate , 2-propanol ,
- Applications
- Analysis of lenampicillin hydrochloride using Shinwasorb
- Column
- Quadrol 5% Shinwasorb-S 60/80
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 100℃
- Carrier gas
- 70 mL/min N₂ (Adjust so that the retention time of cyclohexane is about 1 minute.)
- Injection port temp.
- 160°C
- Detector temp.
- 160°C
- Detector
- FID
- Injection volume
- 4.0 μL
View detail
GC
- Samples
- Enflurane , 2-propanol ,
- Applications
- Analysis of enflurane using Shinwasorb
- Column
- DEGS 10% Shinwasorb-S 60/80
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 80℃
- Carrier gas
- 32 mL/min He (Adjust so that the retention time of enflurane is about 3 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 5.0 μL
View detail
GC
- Samples
- Isoflurane , Ethyl acetate
- Applications
- Analysis of isoflurane using Shinwasorb
- Column
- Igepal CO-880+Ucon LB-550X 5+7.5% Shinwasorb-S 100/120
- Column Size
- 3.5 m x 3.0 mm I.D. Stainless steel
- Column temp.
- 80℃
- Carrier gas
- 37 mL/min N2 (Adjust so that the retention time of isoflurane is about 7 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 2.0 μL
View detail
GC
- Samples
- water, acetonitrile in methanol
- Applications
- Analysis of panipenem using Sunpak-A ~Water~
- Column
- Sunpak-A 80/100
- Column Size
- 2.1m x 3.2 mm I.D. Glass
- Column temp.
- 125℃
- Carrier gas
- 29 mL/min He (Adjust so that the retention time of acetonitrile is about 8 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Cyclohexane , Dicyclohexyl phthalate , Etofenprox
- Applications
- Analysis of etofenprox using Shinwasorb
- Column
- Silicone OV-225 2.5% Shinwasorb-U 60/80
- Column Size
- 1.1m x 3.2 mm I.D. Glass
- Column temp.
- 240℃
- Carrier gas
- 37 mL/min N₂ (Adjust so that the retention time of dicyclohexyl phthalate is about 8 minutes.)
- Injection port temp.
- 250°C
- Detector temp.
- 250°C
- Detector
- FID
- Injection volume
- 3.0 μL
View detail
GC
- Samples
- o-Dichlorobenzene, m-Bromochlorobenzene in acetone
- Applications
- Analysis of o-dichlorobenzene using Shinwasorb
- Column
- PEG20M-TPA 2.5% Shinwasorb-U 60/80
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 115℃
- Carrier gas
- 39 mL/min N₂ (Adjust so that the retention time of o-dichlorobenzene is about 11 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 2.0 μL
View detail